Vötsch
Type: VC 7330/S
Climatic chamber water-cooled
Room volume 3300 litres
Data for temperature checks:
Temperature range -70 °C to + 150 °C
Temperature deviation spatially ± 1 to ± 2 K (corresponds to
temperature gradient) (2 to 4)
Temperature fluctuation in the middle of the usable space ± 0.8 - 1
K
Temperature change speed (with test material: approx. 20 kg
aluminium/plastic)
Heating in the range -40 °C to +90 °C 4 K/min linear
Cooling in the range +90 °C to -40 °C 4 K/min linear
Data for climate testing:
Temperature range +10 to +95°C
Humidity range 10%RH to 95%RH
Temperature fluctuation over time ± 0.2 to - 0.5 K
Humidity fluctuation temporal ± 1%RH up ± 3 %RH
Dew point range +5°C to +88°C
Data for mechanical stress:
of the test room floor 150 kg
max. chamber floor load 250 kg
Dimensions of the test room: width 2000 mm x depth 1150 mm x height
1510 mm
External dimensions: width 3500 mm x depth 1450 mm x height 2100 mm
Data for operation:
Test Room Lighting Halogen Bulb 12 V, 20W
Rated voltage: 3/N/PE AC 400 V ± 10% 50Hz
Rated power 32 kW Rated current 59 A
Of course, external control via the RS 232 interface is possible.
Refrigerant: R404A/R23
Weight of test chamber approx. 800 kg
Weight of machine set approx. 850 kg
Weight: 1650 kg
For you as a buyer A5556157 the following information to be on the
safe side!
The following points are explained in advance at our offered
chambers:
1. Functional check and replacement of necessary components
2. If necessary, refill with legally compliant refrigerants
3. Leak test with certificate
4. After successful verification, the chambers are subjected to a
documented test run.
Condition: gebraucht / used
Scope of delivery: (see picture)
(Changes and errors in the technical data, information is
reserved!)
We will be happy to answer any further questions you may have on the
phone.