The AUTO 3 PTZ decay test instruments are equipped in full compliance
with the requirements of the USP <701/2040> and EP
<2.9.1/2.9.2.1> pharmacopoeias. A large 320x240 pixel LCD
display and alphanumeric keypad allow product and test information to
be changed, such as the decay time for each station, bath and medium
temperature. The total test time for the station is displayed at the
end of a run. The operator intervention to stop a run is also
displayed. Print the individual decay time of each sample when you
have visually detected the disintegration of the sample under test by
simply pressing the corresponding character key (1-6). All this
information can be printed as a test report when a PCL printer is
connected to the parallel printer port of the instruments. Built-in
heating system uses a quiet circulation pump and heating element to
preheat
the outdoor water bath over a range of 25 to 45 °C ; the actual
temperature of the bath is always displayed and
measured by an external manual temperature sensor. Overheat protection
ensures
automatic shut-off of the pump and heating element in the event of no
water flow detection. A
Built-in safety system including overheat detection sensor,
thermoswitch and
Thermal fuse shuts off the heating when the temperature exceeds +60
°C or 95 °C max.
heating element level.